4-Axis setup
Sample tranfer position RT
Please note that top position (T) is not possible to scan all range in y. Follwing values are at room tempereture
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Transfer Samples in/out |
76 |
T: 380 |
25 |
195 |
Screws tight/loose |
76 |
T: 380 |
25 |
105 |
Sample transfer positions for the low temperature head at 10 K
This information seems old, it will be updated.
Positions |
Low temperature head |
multi slot head |
azimuth head |
|---|---|---|---|
a_mp1_x |
66.65 mm |
||
a_mp1_y |
374 mm |
||
a_mp1_z |
17.5 mm |
||
a_mp1_yaw |
2 deg (tight/loose), 357 deg (to insert / remove sample plate) |
Sample position
45deg
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Sample |
74-83 |
T: 23.7 - 35 |
around 22 |
240 |
Focus finder |
78.5 |
97.4 - 97.32 |
23.9 |
240 |
SiN reflection |
78.5 |
97.32 |
23.9 |
240 |
Normal incidende
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Sample |
70-83 |
T: 23.7 - 35 |
? |
283 |
Focus finder |
? |
97.4 - 97.32 |
? |
283 |
SiN reflection |
? |
97.32 |
? |
283 |
6-Axis setup (somebody call 2 axis)
Follwing values are at room tempereture
Sample tranfer position
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
a_mp1_azim |
|---|---|---|---|---|---|
Transfer Samples in/out |
80 |
380 |
15 |
36 |
0 |
Screws tight/loose |
80 |
380 |
15 |
53 |
0 |
Sample position
Normal incidende
2025-11-18 update
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Sample |
80 |
25 |
15 |
306 |
Focus finder |
76.4 - 78.5 |
82.8- 83.980 |
15 |
306 |
45deg
2025-09-19
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Sample |
31 |
25 |
18 |
260 |
Focus finder |
76-75 |
83 |
18 |
260 |
Grazing incidende
posiitons |
a_mp1_x |
a_mp1_y |
a_mp1_z |
a_mp1_yaw |
|---|---|---|---|---|
Sample |
80 |
380 |
15 |
216 |
Focus finder |
80 |
380 |
15 |
216 |
SiN reflection |
80 |
380 |
15 |
216 |
If Manipulator Motor axis is not moving at all
First, please check it is not physically hit something. Generaly, this should not be the case unless the hardware limit switch is hit.
If the hardware limit switch is nit hit, yet the motor not moving then try esync it. In terminal, run ‘spock’. run ‘ipap_esync’ followed by the name of axis: a_mp1_x, a_mp1_y, a_mp1_z, a_mp1_yaw. Eg.: to esync ‘a_mp1_x’ axis, run ‘ipap_esync a_mp1_x’
If the motor is still not moving, reset icepap by enable/disable “pysical” switch. hold toggle to disable until LED indicator change (around 2sec). Hold toggle switch to enable side for around 4 sec. and if status LED indicator becomes green it is recoved. This was reqired once for a_mp1_y axis which has brake on it.
If the problem persists, reach out to the floor coordinator or the local contact.
Note: this axis is not connected to PLC, so emergency stop for q-angle does’t stop this axis.