4-Axis setup

Sample tranfer position RT

Please note that top position (T) is not possible to scan all range in y. Follwing values are at room tempereture

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Transfer Samples in/out

76

T: 380
M: 400
B: 419

25

195

Screws tight/loose

76

T: 380
M: 400
B: 419

25

105

Sample transfer positions for the low temperature head at 10 K

This information seems old, it will be updated.

Positions

Low temperature head

multi slot head

azimuth head

a_mp1_x

66.65 mm

a_mp1_y

374 mm

a_mp1_z

17.5 mm

a_mp1_yaw

2 deg (tight/loose), 357 deg (to insert / remove sample plate)

Sample position

45deg

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Sample

74-83

T: 23.7 - 35
M: 43 - 58
B: 66 - 81

around 22

240

Focus finder

78.5

97.4 - 97.32

23.9

240

SiN reflection

78.5

97.32

23.9

240

Normal incidende

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Sample

70-83

T: 23.7 - 35
M: 43 - 58
B: 66 - 81

?

283

Focus finder

?

97.4 - 97.32

?

283

SiN reflection

?

97.32

?

283

6-Axis setup (somebody call 2 axis)

Follwing values are at room tempereture

Sample tranfer position

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

a_mp1_azim

Transfer Samples in/out

80

380

15

36

0

Screws tight/loose

80

380

15

53

0

Sample position

Normal incidende

2025-11-18 update

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Sample

80

25

15

306

Focus finder

76.4 - 78.5

82.8- 83.980

15

306

45deg

2025-09-19

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Sample

31

25

18

260

Focus finder

76-75

83

18

260

Grazing incidende

posiitons

a_mp1_x

a_mp1_y

a_mp1_z

a_mp1_yaw

Sample

80

380

15

216

Focus finder

80

380

15

216

SiN reflection

80

380

15

216

If Manipulator Motor axis is not moving at all

  1. First, please check it is not physically hit something. Generaly, this should not be the case unless the hardware limit switch is hit.

  2. If the hardware limit switch is nit hit, yet the motor not moving then try esync it. In terminal, run ‘spock’. run ‘ipap_esync’ followed by the name of axis: a_mp1_x, a_mp1_y, a_mp1_z, a_mp1_yaw. Eg.: to esync ‘a_mp1_x’ axis, run ‘ipap_esync a_mp1_x’

  3. If the motor is still not moving, reset icepap by enable/disable “pysical” switch. hold toggle to disable until LED indicator change (around 2sec). Hold toggle switch to enable side for around 4 sec. and if status LED indicator becomes green it is recoved. This was reqired once for a_mp1_y axis which has brake on it.

If the problem persists, reach out to the floor coordinator or the local contact.

Note: this axis is not connected to PLC, so emergency stop for q-angle does’t stop this axis.